PalmSens handheld potentiostat / galvanostat

The opto-isolated cable as well as the Bluetooth option allows PalmSens to be used floating with respect to ground.
PalmSens is delivered with a shielded cable with connections for a working, a counter and a reference electrode by means of 2 mm bananaconnectors with mini-crocodile clips.
The supplied testsensor is used to test the performance of the instrument.
PalmSens can be extended with the bipotentiostat module as well as a multiplexer with 8 or 16 channels.
The PalmSens and bipotentiostat boards are available as OEM potentiostat for other companies for instance for use in an SECM or SPR instrument.
Controlled potential mode or potentiostat
| - dc-potential range | ± 2 V |
| - compliance voltage | ± 8 V |
| - dc-potential resolution | 1 mV |
| - potential accuracy | ≤ 0.2 % offset and 0.2 % |
| - ac-potential amplitude | 1 mV to 250 mV |
| - ac-potential resolution | 0.25 mV |
| - current ranges | 1 nA to 10 mA (8 ranges) |
| - maximum current | ± 10 mA |
| - current resolution | 0.1 % of current range 1 pA on lowest current range |
| - accuracy | ≤ 0.2 % of current range at 100 |
Controlled current mode or galvanostat
| - current ranges | 1 μA to 1 mA |
| - dc-current range | -2 to +2 times selected current range |
| - dc-current resolution | 0.1 % of selected current range |
| - current accuracy | ≤ 0.2 % offset and 0.4 % |
| - maximum output voltage | ± 8 V |
General
| - electrometer amplifier input | > 100 Gohm // 4 pF |
| - rise time | < 50 μs |
| - dimensions | 155 mm x 85 mm x 35 mm |
| - weight | 0.45 kg |
| - power | 2 AA cells NiMH for more than 8 hours operation. battery charger included (6 V- 1500 mA) |
| - interfacing | RS-232 or USB (optionally opto-isolated) or Bluetooth |
* Minimum requirements for netbook, laptop or desktop PC: Windows XP or Vista
* Minimum requirements for Pocket PC: Windows Mobile and serial CF Card