Description
Check if your device is doing what it should with the three circuits of this dummy cell.
A standard dummy cell, that is delivered with PalmSens instruments which include a cell cable. This dummy cell has 3 different circuits that can be used to check the condition of the instrument:
-
- Circuit with two diodes simulating a non diffusion limited redox system. Used to evaluate the performance for low current measurements.
- A high precision 10 kΩ resistor. Recommended for DC measurements.
- A simplified Randle’s circuit (no Warburg impedance). Recommended for EIS experiments.
Resistors are ±1% error, unless indicated. Capacitors are ±5% error.
Specifications
-
-
-
EmStat Pico Development Kit, EmStat3, EmStat3 4WE, EmStat3 Blue, EmStat3+, EmStat3+ Blue, EmStat4M HR Development Kit, EmStat4M LR Development Kit, EmStat4R, EmStat4S HR, EmStat4S LR, EmStat4X HR, EmStat4X LR, MultiEmStat3, MultiEmStat3+, MultiEmStat4 HR, MultiEmStat4 LR, MultiPalmSens4, PalmSens2, PalmSens3, PalmSens4, Sensit BT